Saturday, December 22, 2018
Thursday, December 20, 2018
20181211-12 Quarterly Calibration
Installed vacuum pump on NOy instrument because new one that was installed ceased. (Sergio)
Performed Quarterly Calibration for O3, and CO on 12/11/18 and NOx, NOy on 12/12/18.
Anemometer needs to be rebuild (Sergio)
Performed Quarterly Calibration for O3, and CO on 12/11/18 and NOx, NOy on 12/12/18.
Launch
Trailer Slopes
O3 0.98 (prev. 1.026)
CO 0.85 (prev.
0.85)
NO 0.447 (prev. 0.41)
NOx 0.95 (prev. 0.98)
NOy 1.090 (prev.
1.097)
NOx CE 72.276 (prev. 73.122)
Slope 0.75 (prev. 0.75)
Slope 0.75 (prev. 0.75)
NOy CE 90.73 (prev. 97.93)
Slope 0.91 (Prev. 0.98)
Slope 0.91 (Prev. 0.98)
Anemometer needs to be rebuild (Sergio)
Thursday, December 6, 2018
Monday, October 29, 2018
Monday, October 15, 2018
Monday, September 17, 2018
20180917 LT quarterly calibratio
CO,O3,NOx, NOy quarterly calibration
replaced Teflon filters
CO, 1800
NO: 500
NO2: 1700
replaced Teflon filters
CO, 1800
NO: 500
NO2: 1700
Tuesday, September 11, 2018
20180911 LT Ozone instrument turned on. (additional notes within)
- Ozone instrument stopped streaming data on Sunday, 9/9/2018 because it turned "itself" off. On Monday morning the power was cycled on and off, and it turned on.
Monday, August 6, 2018
Sunday, August 5, 2018
20180806 LT CO sample flow very low
Sample flow decreased substantially on Saturday afternoon about 6 PM CST, so the instrument was set to "maintenance mode."
Wednesday, August 1, 2018
20180730 LT Performed unscheduled multipoint calibration
Note: NOxy instruments were multipoint calibrated because the NO instrument spans decreased. The slopes will calculated and compared to the latest calibration and adjustment will be made accordingly.
Thursday, July 19, 2018
20180719 LT site visit
Alarms:
42C (top) NO-NO2-NOx alarm:
1. Converter temp low, 0 C/300 C minimum
42i-TL NO-NO2-NOx alarms:
1. Low pressure: 147 mmHg/200 mmHg minimum
2. Low flow: 0.702 L/min / 0.750 L/min minimum
42C (bottom) NO-NO2-NOx alarms:
1. Converter temp low, 0 C/300 C minimum
2. Sample flow low: 0.156 LPM/ 0.200 LMP minimum
Cylinder pressures:
CO2: 2.076%, 2000 psig
NO: 4.95 ppm, 500 psig
CO: 25.90 ppm, 1750 psig
NO2: 4.91ppm, 1740 psig
-Changed particle sample filers
-Cleaned cooling fan filters
-Checked the compressor drain
-Checked anemometer for bad bearings
-Checked vacuum pumps for chafing tubing
-Checked that molybdenum NOy converter temp is holding at 320 C
-Replaced IR sensor in CO instrument
-Replaced O3 flow sensor
42C (top) NO-NO2-NOx alarm:
1. Converter temp low, 0 C/300 C minimum
42i-TL NO-NO2-NOx alarms:
1. Low pressure: 147 mmHg/200 mmHg minimum
2. Low flow: 0.702 L/min / 0.750 L/min minimum
42C (bottom) NO-NO2-NOx alarms:
1. Converter temp low, 0 C/300 C minimum
2. Sample flow low: 0.156 LPM/ 0.200 LMP minimum
Cylinder pressures:
CO2: 2.076%, 2000 psig
NO: 4.95 ppm, 500 psig
CO: 25.90 ppm, 1750 psig
NO2: 4.91ppm, 1740 psig
-Changed particle sample filers
-Cleaned cooling fan filters
-Checked the compressor drain
-Checked anemometer for bad bearings
-Checked vacuum pumps for chafing tubing
-Checked that molybdenum NOy converter temp is holding at 320 C
-Replaced IR sensor in CO instrument
-Replaced O3 flow sensor
Thursday, July 12, 2018
20180711 Ozone instrument was turned back on about noon because instrument power was off.
Ozone data stopped logging about 4 am cst
Thursday, June 21, 2018
20180530 Quarterly calibration
Quarterly calibrations were performed
Teflon filters were replaced
Rain gauge calibration performed
Teflon filters were replaced
Rain gauge calibration performed
Monday, April 30, 2018
20180420 Site Check
- changed sample particle filters on NO, NOx, NOy, CO, and Ozone instruments
- instrument alarms:
NO instrument - converter temperature - NA
NOx instrument - flow low alarm - 0.459/0.805 LPM
pressure low alarm - 190/148 mm Hg
NOy instrument - converter temperature high -NA
-sample flow mow - 0 LPM (sensor needs to be replaced)
Calibrator 146C - No alarms
O3 instrument - flow B high - 1.5 LPM (sensor will be replaced)
CO instrument - no alarms
- Cylinder pressures - NO2 cylinder CC99506 1750 psig
NO cylinder CB08891 600 psig
CO cylinder CA08706 (?) psig
- instrument alarms:
NO instrument - converter temperature - NA
NOx instrument - flow low alarm - 0.459/0.805 LPM
pressure low alarm - 190/148 mm Hg
NOy instrument - converter temperature high -NA
-sample flow mow - 0 LPM (sensor needs to be replaced)
Calibrator 146C - No alarms
O3 instrument - flow B high - 1.5 LPM (sensor will be replaced)
CO instrument - no alarms
- Cylinder pressures - NO2 cylinder CC99506 1750 psig
NO cylinder CB08891 600 psig
CO cylinder CA08706 (?) psig
Friday, April 6, 2018
Tuesday, April 3, 2018
Monday, April 2, 2018
Friday, March 30, 2018
20180330 Connected zero air from Aadco zero air system to ozone instrument zero air port
- Two Google Earth cars with NO, NO2, CO, Blach carbon and PM 2.5 parked next to site for intercomparison
Monday, March 12, 2018
20180312 Anemometer was replaced with Model 05305-AQ
- new data logger program was uploaded with appropriate propellor coefficient