20191018 Monthly Check
No leaks or smells
Pumps clear
No water in charcoal canisters
No condensation in the sample lines
Changed Teflon Filters
NO flow: >1
NO pressure: 250
NO PMT temp:
NO alarms: converter temp
NOx Flow: 0.5-.75
NOx pressure: 125-250
NOx PMT temperature:
NOx alarms: converter temp
O3 flow: A:1.51, B:0.70
O3 pressure: 727
O3 alarms: FLow A high, intensity b low
NO2 flow: 1.25-1.50
NO2 pressure: 250
NO2 PMT temp:
NO2 alarms Cooler temp high, converter temp low, sample flow high
CO flow: 0.50
CO pressure: 741
CO alarms: none
NO high: 200 NO low: 42
NO2 high: 1600 NO2 low: 34
Co2 High: 1250 CO2 low: 52
CO high: 1700 CO low: 40
No pests
Pumps clear
No water in charcoal canisters
No condensation in the sample lines
Changed Teflon Filters
NO flow: >1
NO pressure: 250
NO PMT temp:
NO alarms: converter temp
NOx Flow: 0.5-.75
NOx pressure: 125-250
NOx PMT temperature:
NOx alarms: converter temp
O3 flow: A:1.51, B:0.70
O3 pressure: 727
O3 alarms: FLow A high, intensity b low
NO2 flow: 1.25-1.50
NO2 pressure: 250
NO2 PMT temp:
NO2 alarms Cooler temp high, converter temp low, sample flow high
CO flow: 0.50
CO pressure: 741
CO alarms: none
NO high: 200 NO low: 42
NO2 high: 1600 NO2 low: 34
Co2 High: 1250 CO2 low: 52
CO high: 1700 CO low: 40
No pests
Wednesday, October 16, 2019
20191016 CO Diaphragm Replaced
CO flow was low during morning check.
Replaced diaphragm with a used one from 403 because we are waiting on more to be shipped
Replaced diaphragm with a used one from 403 because we are waiting on more to be shipped
Monday, October 14, 2019
20191011 Ozone instrument sample line fitting
Ozone was not reading normally, disconnected and reconnected sample line resolved issue, instrument had been sampling inside the trailer (60ppb).
20191010 Ozone pump failure.
On arrival temp inside trailer was very cold, temp was at 67F set to 60F. Terailer smelled heavily of Ozone. Door was opened and AC was set to 80F. (Eugenia) Replaced Ozone instrument diaphragm, instrument was turned back online.
Monday, September 30, 2019
20190930 CO Fans Replaced
CO instrument was shut down at 12:13 CST to replace bench fan, and exhaust fan.
Instrument was restarted at 13:01 CST. New fans are slightly lower wattage, they no longer had same model of fans for sale online. Wires for the fans had to be clipped to remove connector, and were connected to fans using screw terminal blocks.
Friday, September 27, 2019
20190924 NOy Pressure Spike
Over the weekend NOy pressure gradually increased to ambient pressure until Sunday when pressure suddenly dropped back to normal. Tuesday Daniel Bahrt went on site to investigate potential cause. The exhaust line from the NOy pump had some droplets of hardened residue inside and looked like it had been blocked off then blown back out. Ozone Catalyst at the end of the pump may need to be replaced.
Tuesday, September 24, 2019
20190924 Site Check
Post tropical storm check
No signs of water damage/leaks
NO-less than 200
NO2-1600
CO- 1700
Anemometer working
No signs of water damage/leaks
NO-less than 200
NO2-1600
CO- 1700
Anemometer working
Thursday, September 5, 2019
20190903 CO Fan failure
Started receiving high temperature alerts, trailer temp was fine, co instrument had temp bouncing from 35C to 45C. Upon opening the instrument all 3 fans were not working properly. The bench fan was spinning slowly (like a stepper motor) and was visually cracked open. The exhaust fan was also clicking as it slowly ran. All 3 fans were hot and the instrument was warm to the touch. The instrument was shut down and restarted at 15:22CST to test. Bench fan was determined to be dead and was replaced by a fan from an O3 analyzer instrument located in 403. 2 Fans need to be ordered and installed to replace the clicking exhaust fan and the bench fan. The instrument has not had all fan screws reinstalled, but was restarted in maintenance mode.
Tuesday, September 3, 2019
20190902 AC Outage
High instrument temperatures indicated possible AC outage. When Daniel arrived onsite temperature inside the trailer was 114F and AC was not running. The door was opened to let the air cool and HVAC was called. HVAC was able to arrive that day so it was decided not to shut down the site. HVAC said that they replaced an alternator. Daniel stopped by again later to verify the temperature had dropped. Inside temp was 94F and cool (not cold) air was blowing from the vents.
Monday, August 26, 2019
20190808 Computer Issue
8/7/2019 Computer screen was not working properly. Brought new VGA cable that fixed the problem
Wednesday, August 7, 2019
20190807 Site Visit
No pests/leaks
NO- 250 psi NO2-1600 psi CO-1600 psi
CO-no alarms; Dynamic Gas-no alarms; NO-converter temp low;NOy-converter temp low
Changed teflon filters
NO- 250 psi NO2-1600 psi CO-1600 psi
CO-no alarms; Dynamic Gas-no alarms; NO-converter temp low;NOy-converter temp low
Changed teflon filters
Friday, July 19, 2019
20190718 NO Zero air Valve troubleshoot
Zero air valve had previously been connected to the AUX connection on the motherboard, this was wrong and moved to the SAMPLE connection after verifying with multimeter. Daqfactory did not have proper sequences or commands to zero NO. The sequence "Auto_Zero_NO" was re-written with proper commands for the instrument and was turned on. The sequence is not currently changing the baseline, and that will have to be written in there when next available.
Tuesday, June 11, 2019
20190611 Labjack Not Found
Labjack not found error identified by Sergio as potential cause of no spans
USB hub found to be malfunctioning (not all green lights on)
USB hub power cycled, seems to have fixed error.
**NOTE- USB hub was warm to the touch**
USB hub found to be malfunctioning (not all green lights on)
USB hub power cycled, seems to have fixed error.
**NOTE- USB hub was warm to the touch**
Tuesday, May 28, 2019
Thursday, May 9, 2019
20190509 Site check (refer to notes)
- changed filters on NO, NOx, NOy, CO, and Ozone
- checked NOy converter - 320 deg C
- checked instrument alarms: Usual not applicable alarms were seen except for the Ozone instrument had a low intensity alarm
- checked NOy converter - 320 deg C
- checked instrument alarms: Usual not applicable alarms were seen except for the Ozone instrument had a low intensity alarm
Wednesday, May 8, 2019
Monday, March 18, 2019
20190315 LT quarterly calibrations
Performed quarterly calibrations for CO, O3, NO, NOx, NOy
Replaced Teflon filters
Replaced Teflon filters
20190306 Restarted site and did maintenance
- Replaced IR source and drying tube on the CO instrument
- Turned O3, CO, NO, NOy, and calibrator on in preparation for the Ozone measuring season.
- Replaced Vacumbrand pump
- Unplugged BIC plug in the inlet box
- CO= 1640 psi
- NO= 350 psi